Protein Sample Concentration: Difference between revisions

From Microbial Ecology and Evolution Lab Wiki
Jump to navigation Jump to search
PreDec2022>TrevorE
m (10 revisions imported)
 
(One intermediate revision by one other user not shown)
(No difference)

Latest revision as of 12:57, 16 December 2022

Goal

Use the Amicon® Ultra-4 centrifugal filter device to concentrate a protein sample through rapid ultrafiltration. The following information and protocol was adapted from here

Before You Start

Picking The Right Filtering Device

The membranes used in Amicon® Ultra devices are characterized by a molecular weight cutoff (MWCO); that is, their ability to retain molecules above a specified molecular weight. Solutes with molecular weights close to the MWCO may be only partially retained. For best results, use a membrane with a MWCO at least two times smaller than the molecular weight of the protein solute that one intends to concentrate. See table below (Amicon® User Guide Table 4):

Table 4 - User Guide.png


What You Will Need

  • Sample to concentrate
  • Swinging-bucket centrifuge (Bio Superlab)
  • Conical-bottomed tubes
  • P200 or P1000 Pipettor with appropriate tips
  • Appropriate MWCO Amicon® Ultra filter device (Amicon® User Guide Figure 1)

Figure 1 - Amicon® Ultra-4 device.png


Tips

Maximize Sample Recovery

Low sample recovery in the concentrate may be due to adsorptive losses, over-concentration, or passage of sample through the membrane.

  1. Adsorptive losses depend upon solute concentration, its hydrophobic nature, temperature and time of contact with filter device surfaces, sample composition, and pH. To minimize losses, remove concentrated samples immediately after centrifugal spin.
  2. Over-concentration can lead to precipitation and potential sample loss. If the starting sample concentration is high, monitor the centrifugation process in order to avoid over-concentration of the sample.'
  3. If the sample appears to be passing through the membrane, choose a lower MWCO Amicon® Ultra-4 device.


Desalting or Diafiltration

Desalting, buffer exchange, or diafiltration are important methods for removing salts or solvents in solutions containing biomolecules.

They can be accomplished by concentrating the sample, then reconstituting the concentrate to the original sample volume with any desired solvent. The process of “washing out” can be repeated until the concentration of the contaminating microsolute has been sufficiently reduced. See process overview below (Amicon® user guide Figure 2).

Desalitnation Process - User Guide Amicon Ultra-4 Centrifugal Filter Devices.png

Procedure

  1. Rinse Amicon® Ultra filter device with buffer or Milli-Q® water before use
    • CAUTION: Do not allow the membrane in Amicon® Ultra filter devices to dry out once wet. If you are not using the device immediately after rinsing, leave fluid on the membrane until the device is used.
  2. Add up to 4 mL of sample (3.5 mL if using a 23° fixed-angle rotor) to the Amicon® Ultra filter device.
  3. Place capped filter device into centrifuge rotor
  4. Centrifuge setting options:
    1. Swinging-bucket (preferred); counterbalance with a similar device and water placed in the centrifuge tube (not the balance filter). Ensure weight of both centrifuge tubes and filters are within 0.1g of each other. Spin at 4,000 × g maximum for approximately 10–40 minutes.
  5. To recover the concentrated solute, insert a pipettor into the bottom of the filter device and withdraw the sample using a side-to-side sweeping motion to ensure total recovery. The ultrafiltrate can be stored in the centrifuge tube.
    NOTE: For optimal recovery, remove concentrated sample immediately after centrifugation.

Table 2 and 3 - User Guide Amicon Ultra-4 Centrifugal Filter Devices.png